Benchtop Multimeter Keithley 2000, 6-1/2-Digit, GPIB/RS232
Hotline: +84 906 988 447
Head Office: Ho Chi Minh City
- Tel: +84 2839 778 269 / 3601 6797
- Email: [email protected]
- Add: 487 Cong Hoa Street, Tan Binh Ward, Ho Chi Minh City, Vietnam
Office: Bac Ninh City
- Tel: +84 222 730 0180
- Email: [email protected]
- Add: 184 Binh Than Street, Vo Cuong Ward, Bac Ninh, Vietnam
-
Technical Counseling
100% Free
-
Free Shipping
For 3.000.000vnd Order
Data is being updated
Description
The 6½-digit 2000 Series offers a unique combination of high-precision, cost-effectiveness, and flexibility. Based on the same high-speed, low-noise A/D converter technology found in our 7½- and 8½-digit multimeters, this DMM offers market-leading speed and accuracy. For greater flexibility, a built-in card slot on the rear panel allows an additional multiplexer card for multi-point measurement applications.
Features
- 13 built-in measurement functions
- 2000 readings/second at 4-1/2 digits
- Optional scanner cards for multipoint measurements
- GPIB and RS-232 interfaces
- Fluke 8840/42 command set
| General Specification | ||
| POWER SUPPLY | 100V / 120V / 220V / 240V | |
| LINE FREQUENCY | 50Hz to 60Hz and 400Hz, automatically sensed at power-up | |
| POWER CONSUMPTION | 22 VA | |
| OPERATING ENVIRONMENT | Specified for 0°C to 50°C. Specified to 80% R.H. at 35°C. Altitude up to 2000 meters. | |
| STORAGE ENVIRONMENT | –40°C to 70°C | |
| WARRANTY | 1 year | |
| EMC | Conforms to European Union EMC Directive | |
| SAFETY | Conforms to European Union Low Voltage Directive | |
| VIBRATION | MIL-PRF-28800F Class 3 Random | |
| WARMUP | 1 hour to rated accuracy | |
| DIMENSIONS | Rack Mounting | 89mm high x 213mm wide x 370mm deep (3.5 in. x 8.38 in. x 14.56 in.) |
| Bench Configuration (with handle and feet) | 104mm high x 238mm wide x 370mm deep (4.13 in. x 9.38 in. x 14.56 in.) | |
| NET WEIGHT | 2.9kg (6.3 lbs) | |
| SHIPPING WEIGHT | 5kg (11 lbs) | |
| VOLT HERTZ PRODUCT | =8 x 10 7 V·Hz | |
| DC CHARACTERISTICS | ||||||||
| CONDITIONS | MED (1 PLC) 1 or SLOW (10 PLC) or MED (1 PLC) with filter of 10 | |||||||
| ACCURACY | ±(ppm of reading + ppm of range) (ppm = parts per million) (e.g., 10ppm = 0.001%) | |||||||
| FUNCTION | RANGE | RESOLUTION | TEST CURRENT OR BURDEN VOLTAGE (±5%) | INPUT RESISTANCE | 24 HOUR 23°C ± 1° | 90 DAY 23°C ± 5° | 1 YEAR 23°C ± 5° | TEMPE- RATURE COEFF- ICIENT 0°–18°C & 28°–50°C |
| Voltage | 100.0000mV | 0.1 µV | >10GΩ | 30+30 | 40+35 | 50+35 | 2+6 | |
| 1.000000V | 1.0µV | >10GΩ | 15+6 | 25+7 | 30+7 | 2+1 | ||
| 10.00000V | 10µV | >10GΩ | 15+4 | 20+5 | 30+5 | 2+1 | ||
| 100.0000V | 100µV | 10MΩ±1% | 15+6 | 30+6 | 45+6 | 5+1 | ||
| 1000.000V | 1mV | 10MΩ±1% | 20+6 | 35+6 | 45+6 | 5+1 | ||
| Resistance | 100.0000Ω | 100µΩ | 1mA | 30+30 | 80+40 | 100+40 | 8+6 | |
| 1.000000kΩ | 1mΩ | 1mA | 20+6 | 80+10 | 100+10 | 8+1 | ||
| 10.00000kΩ | 10mΩ | 100µA | 20+6 | 80+10 | 100+10 | 8+1 | ||
| 100.0000kΩ | 100mΩ | 10µA | 20+6 | 80+10 | 100+10 | 8+1 | ||
| 1.000000MΩ | 1Ω | 10µA | 20+6 | 80+10 | 100+10 | 8+1 | ||
| 10.00000MΩ | 10Ω | 700nA// 10MΩ | 150+6 | 200+10 | 400+10 | 95+1 | ||
| 100.0000MΩ | 100Ω | 700nA// 10MΩ | 800+30 | 1500+30 | 1500+30 | 900+1 | ||
| Current | 10.00000mA | 10nA | <0.15V | 60+30 | 300+80 | 500+80 | 50+5 | |
| 100.0000mA | 100nA | <0.03V | 100+300 | 300+800 | 500+800 | 50+50 | ||
| 1.000000A | 1µA | <0.3V | 200+30 | 500+80 | 800+80 | 50+5 | ||
| 3.00000A | 10µA | <1V | 1000+15 | 1200+40 | 1200+40 | 50+5 | ||
| Current | 1 kΩ | 100 mΩ | 1 mA | 40+100 | 100+100 | 120+100 | 8+1 | |
| Diode Test | 3.00000V | 10 µV | 1 mA | 20+6 | 30+7 | 40+7 | 8+1 | |
| 10.00000V | 10 µV | 100µA | 20+6 | 30+7 | 40+7 | 8+1 | ||
| 10.00000V | 10 µV | 10µA | 20+6 | 30+7 | 40+7 | 8+1 | ||
| DC OPERATING CHARACTERISTICS | ||||||||
| FUNCTION | DIGITS | READINGS/s | PLCs | |||||
| DCV (all ranges), DCI (all ranges), and Ohms (<10M range) | 6.5 | 5 | 10 | |||||
| 6.5 | 30 | 1 | ||||||
| 6.5 | 50 | 1 | ||||||
| 5.5 | 270 | 0.1 | ||||||
| 5.5 | 500 | 0.1 | ||||||
| 5.5 | 1000 | 0.04 | ||||||
| 4.5 | 2000 | 0.01 | ||||||
| DC SYSTEM SPEEDS | ||||||||
| RANGE CHANGE | 50/s | |||||||
| FUNCTION CHANGE | 45/s | |||||||
| AUTORANGE TIME | <30 ms | |||||||
| ASCII READINGS TO RS-232 (19.2K BAUD) | 55/s | |||||||
| MAX. INTERNAL TRIGGER RATE | 2000/s. | |||||||
| MAX. EXTERNAL TRIGGER RATE | 500/s | |||||||
| DC GENERAL | ||||||||
| LINEARITY OF 10VDC RANGE | ±(1ppm of reading + 2ppm of range) | |||||||
| DCV, Ω, TEMPERATURE, CONTINUITY, DIODE TEST INPUT PROTECTION | 1000V, all ranges | |||||||
| MAXIMUM 4WΩ LEAD RESISTANCE | 10% of range per lead for 100Ω and 1kΩ ranges; 1kΩ per lead for all other ranges | |||||||
| DC CURRENT INPUT PROTECTION | 3A, 250V fuse | |||||||
| SHUNT RESISTOR | 0.1Ω for 3A, 1A and 100mA ranges. 10Ω for 10mA range | |||||||
| CONTINUITY THRESHOLD | Adjustable 1Ω to 1000Ω | |||||||
| AUTOZERO OFF ERROR | Add ±(2ppm of range error + 5µV) for <10 minutes and ±1°C change | |||||||
| OVERRANGE | 120% of range except on 1000V, 3A and Diode | |||||||
| SPEED AND NOISE REJECTION | ||||||||
| RATE | READINGS/S | DIGITS | RMS NOISE 10V RANGE | NMRR | CMRR | |||
| 10 PLC | 5 | 6.5 | < 1.5 µV | 60 dB | 140 dB | |||
| 1 PLC | 50 | 6.5 | < 4 µV | 60 dB | 140 dB | |||
| 0.1 PLC | 500 | 5.5 | < 22 µV | 80 dB | ||||
| 0.01 PLC | 2000 | 4.5 | < 150 µV | 80 dB | ||||
| DC Notes | ||||||||
| ||||||||
| TRUE RMS AC VOLTAGE AND CURRENT CHARACTERISTICS | |||||||
| ACCURACY | ±(% of reading + % of range), 23°C ±5 °C | ||||||
| VOLTAGE RANGE | RESOLUTION | CALIBRATION CYCLE | 3 Hz– 10 Hz | 10 Hz– 20 kHz | 20 kHz– 50 kHz | 50 kHz– 100 kHz | 100 kHz– 300 kHz |
| 100.0000mV | 0.1 µV | ||||||
| 1.000000V | 1.0 µV | 90 Days | 0.35+0.03 | 0.05+0.03 | 0.11+0.05 | 0.60+0.08 | 4+0.5 |
| 10.00000V | 10 µV | ||||||
| 100.0000V | 100 µV | 1 Year | 0.35+0.03 | 0.06+0.03 | 0.12+0.05 | 0.60+0.08 | 4+0.5 |
| 750.000V | 1 mV | ||||||
| TEMPERATURE COEFFICIENT/°C | 0.035+ 0.003 | 0.005+ 0.003 | 0.006+ 0.005 | 0.01+ 0.006 | 0.03+ 0.01 | ||
| CURRENT RANGE | RESOLUTION | CALIBRATION CYCLE | 3 Hz– 10 Hz | 10 Hz– 3 kHz | 3 kHz- 5 kHz | ||
| 1.000000A | 1 µA | 90 Day/ 1 Year | 0.30+ 0.04 | 0.10+ 0.04 | 0.14+ 0.04 | ||
| 3.00000A | 10 µA | 90 Day/ 1 Year | 0.35+ 0.06 | 0.15+ 0.06 | 0.18+ 0.06 | ||
| TEMPERATURE COEFFICIENT/°C | 0.035+ 0.003 | 0.015+ 0.006 | 0.015+ 0.006 | ||||
| HIGH CREST FACTOR ADDITIONAL ERROR ±(% of reading) | |||||||
| CREST FACTOR | 1–2 | 2–3 | 3–4 | 4–5 | |||
| ADDITIONAL ERROR | 0.05 | 0.15 | 0.30 | 0.40 | |||
| AC OPERATING CHARACTERISTICS | |||||||
| FUNCTION | DIGITS | READINGS/s | RATE | BANDWIDTH | |||
| ACV (all ranges), and ACI (all ranges) | 6.5 | 2s/reading | SLOW | 3 Hz–300 kHz | |||
| 6.5 | 1.4 | MED | 30 Hz–300 kHz | ||||
| 6.5 | 4.8 | MED | 30 Hz–300 kHz | ||||
| 6.5 | 2.2 | FAST | 300 Hz–300 kHz | ||||
| 6.5 | 35 | FAST | 300 Hz–300 kHz | ||||
| ADDITIONAL LOW FREQUENCY ERRORS ±(% of reading) | |||||||
| SLOW | MED | FAST | |||||
| 20Hz – 30Hz | 0 | 0.3 | |||||
| 30Hz – 50Hz | 0 | 0 | |||||
| 50Hz – 100Hz | 0 | 0 | 1.0 | ||||
| 100Hz – 200Hz | 0 | 0 | 0.18 | ||||
| 200Hz – 300Hz | 0 | 0 | 0.10 | ||||
| > 300Hz | 0 | 0 | 0 | ||||
| AC SYSTEM SPEEDS | |||||||
| FUNCTION/RANGE CHANGE | 4/s | ||||||
| AUTORANGE TIME | <3s | ||||||
| ASCII READINGS TO RS-232 (19.2k BAUD) | 50/s | ||||||
| MAX. INTERNAL TRIGGER RATE | 300/s | ||||||
| MAX. EXTERNAL TRIGGER RATE | 300/s | ||||||
| AC GENERAL | |||||||
| INPUT IMPEDANCE | 1MΩ ±2% paralleled by <100pF | ||||||
| ACV INPUT PROTECTION | 1000Vp | ||||||
| MAXIMUM DCV | 400V on any ACV range | ||||||
| ACI INPUT PROTECTION | 3A, 250V fuse | ||||||
| BURDEN VOLTAGE | 1A Range: <0.3V rms. 3A Range: <1V rms | ||||||
| SHUNT RESISTOR | 0.1Ω on all ACI ranges | ||||||
| AC CMRR | >70dB with 1kΩ in LO lead | ||||||
| MAXIMUM CREST FACTOR | 5 at full scale | ||||||
| VOLT HERTZ PRODUCT | =8 x 10 7 V·Hz | ||||||
| OVERRANGE | 120% of range except on 750V and 3A ranges | ||||||
| AC Notes | |||||||
| |||||||
Keithley 2000 Product Specifications Summary: (for full specs please see the full specs tab or the data sheet)
Multimeters Template | |
|---|---|
| Type | Digital |
| Style | Bench |
| Measures AC V | Yes |
| Measures DC V | Yes |
| Measures AC A | Yes |
| Measures DC A | Yes |
| True RMS What's This? | Yes |
Test Equipment General Attributes | |
| Power Supply Voltage | 120V 50/60 Hz |
Accessories
-
Current Range:
-
Voltage Range:
-
Display Resolution:
Digital Multimeters Siglent SDM4065A-SC
-
Current Range:
-
Voltage Range:
-
Display Resolution:
Digital Multimeters Siglent SDM4065A
-
Current Range:
-
Voltage Range:
-
Resistance Range:
-
Display Resolution:
Digital Multimeter Kikusui DME1600
-
Current Range:
-
Voltage Range:
-
Resistance Range:
-
Display Resolution:
Digital Multimeter Kikusui DME1600GC
-
Voltage Source:
-
Current Source:
-
Voltage Resolution:
-
Current Resolution:
High Power SourceMeter Keithley 2657A (1fA / 100nV)
Please login to write review!